The Hg-Monitor 3000, the sequel to Hg-MAK 1200 and Hg-Monitor 2000, is a highly sensitive instrument dedicated for continuous elemental mercury vapour measurements in gases such as air, nitrogen, hydrogen, natural gas etc. It has a wide field of applications including ambient air monitoring, exhaust air monitoring, examination of contaminated sites and process monitoring.
- Continuous on-line operation
- Easy to operate, self-explaining software
- Auto-zero for precise measurement
- Available for tabletop, 19" and mobile use
- Precise measuring 0,1 to 3000 µg/m3 due to non-linear compensation in the high measuring range
- Temperature control for light-source and cuvette
- Unique electrodeless mercury lamp with long lifetime (over 20.000 hours)
- Wide range power supply (90 - 265 VAC/VDC, 12 VDC
- Graphic TTF-display and one-touch keypad
- External Keyboard and VGA monitor, remote control by modem possible
- Approved for EMC and low-voltage directives (CE)
- Main component ("heart") of a complete range of mercury analyzers for gases, liquids, solids
- RS 232, RS485, 4-20 mA analogue and PLC outputs
- Self-diagnosis, status outputs, printer port
Measuring PrincipleThe measuring principle used in the HG - MONITOR 3000 is based on the resonance absorption of mercury atoms at a wavelength of 253.7 nm. It exploits the high vapor pressure of mercury and the fact that except for the noble gases, Mercury is the only element that has monoatomic vapor at room temperature.
FunctionalityThe sample gas is continuously fed via a dust filter into the optical cuvette of the HG - MONITOR 3000. Here an absorption measurement of the radiation emitted by a UV light source (253.7 nm Hg line) is carried out. The intensity of the UV lamp is continuously measured via a reference beam and the measurement is corrected accordingly to compensate for possible fluctuations of the light source.
UV Light SourceThe HG - MONITOR 3000's UV source is a low-pressure mercury discharge lamp with high-frequency excitation. This method yields extreme-narrow-band emission lines, which are congruent with the absorption lines of the mercury atoms. This provides improved detection sensitivity and a lower cross-sensitivity compared to other light sources. The lamp has a much longer service life than other UV lamps. The lamp unit is temperature-stabilized.
Easy OperationThe HG - MONITOR 3000 is micro-processor-controlled. After it is switched on the device automatically calibrates itself and then switches to measuring mode. Inputs can be made via the waterproof membrane keyboard. The following parameters are menu-adjustable:
- Display of ug/m3 or ppb
- Conversion to standard conditions
- Time interval for printer output
- Alarm level and prelim level
- 3 time intervals for computation of mean value
- Measuring range 0...20, 0...100, 0...1000, 0...2000 ug/m3 (ranges are user programmable)
Fields of Application
- Mercury in air
- Mercury in gases such as H2 , N2 and Natural Gas.
- Work-place monitoring
- 2 adjustable TLV (Threshold Limit Value) monitoring
- Measurement of mercury in exhaust air
- Ambient sir screening at ground level
- Detector for amalgam technology
Data Display and OutputThe mercury concentration is displayed on a 4-line display and is also outputted to a printer interface and an RS- 232 interface. The output is adjusted for pressure and temperature. Mean values over programmable periods of time (e.g. half-hour mean values) can also be calculated. The pressure and temperature of the sample gas can be displayed by pressing a button. An analog output signal is also available for connecting to a data acquisition system or a chart recorder.
Self-diagnostics systemThe built-in processor constantly monitors the functions of vital components of the HG - MONITOR 3000. If a fault occurs the user is warned via the display and via a separate output signal.
Technical Specifications
|
Measuring Method
|
Cold Vapor Atomic Absorption
|
|
Wavelength
|
254 nm
|
|
UV Lamp
|
Temperature-controlled high frequency excitation lamp
|
|
Cuvette
|
230 mm, quartz glass
|
|
Measuring Ranges
|
0-20, 0-100, 0-1000, 0-2000 ug/m3 (ranges are user programmable)
|
|
(Gaseous phase)
|
ppb display at choice
|
|
Accuracy
|
± 2 % at 100 ug/m3
|
|
Sensitivity
|
0.1 ug/m3
|
|
Temp/Pressure Compensation
|
as option
|
|
Response Time
|
Selectable, 0 / 1 / 5 / 30 Seconds
|
|
Calibration
|
Factory-calibrated
|
|
Alarm
|
2 levels programmable, over full scale
|
|
Mean Value
|
3 different mean values programmable, selectable between 0-999 min
|
|
Output Signals
|
Analog : 0-10 V (optionally 4...20 mA)
Digital : Centronics and RS 232~
Status : (Alarm and Error Message)
|
|
Display
|
TFT color graphic display with background illumination
|
|
Pump
|
Membrane pump, ~ 1.5 l/min
|
|
Filter
|
Dust filter, easy exchange of cartridges
|
|
Power Supply
|
90 - 240 VAC at 50 / 60 Hz, 12 VDC
|
|
Operating Temperature
|
32 to 113 °F (0 to 45 °C )
|
|
Dimensions
|
(Width x Height x Depth) without handle
18 in x 5.2 in x 13 in (449 mm x 133 mm x 331 mm)
|
|
|
Standard with handle, optional version: 19" chassis unit
|
|
Weight
|
12.5 kg
|