Housing:
|
Instrument built-in in a cabinet of stainless steel, protected against ambient according to IP 55
|
|
Method:
|
UV-photometer with measurement path and reference-beam for the stabilisation of the lamp. Cold vapour atomic fluorescence.
|
|
Cuvette:
|
Quartz, optical path length 230 mm
|
|
Lightsource:
|
High frequency excited Mercury low pressure lamp, stabilised light output, temperature stabilised.
|
|
Ranges:
|
0 ... 500 µg Hg per Litre maximum range. Lowest range 0 ... 10 µg/l (higher concentration requires upstream dilution unit, available as option)
|
|
Resolution:
|
0,1 µg Hg per Litre Sample
|
|
Rise time:
|
0,5 to 2 minutes
|
|
Sampling:
|
Continuous sampling from bypass (maximum pressure on bypass is 500 mm H2O)
|
|
Software:
|
interactive operator prompting (according to EN 1483)
|
|
Input:
|
Moisture-proof membrane keyboard with 5 programmable keys
|
|
Display:
|
TFT-colour-display, graphic representation, back-lighting
|
|
Output:
|
Analogue output: 0 – 10 V (option 4...20 mA), RS 232 for digital data processing and software-update
|
|
Pumps:
|
2 peristaltic pumps with 3 pump-heads each (sample, KMnO4, Hydroxyl-Ammonium-Chloride, SnCl2, blank,
|
|
Reagents
|
Potassium-Permanganate, Hydroxyl-Ammonium-Chloride, Tin(II)Chloride, distilled water, mercury standard solution
|
|
Supply of Regents
|
Continuous transport of the reagents by peristaltic pumps, fasts response of signal due to continuous flow injection of reagents
|
|
Reactor
|
New designed reactor with reduced volume and purge module
|
|
Zeroing
|
Zeroing with built-in zero valve for blank solution or water, includes cleaning of the system
|
|
Calibration
|
Semi-manual calibration with standard solution at 50 % of maximum.
|
|
Fuses:
|
Photometer: T 2 A micro-fuse, Cabinet 16 A automatic cutout
|
|
Operating temperature:
|
5 – 40°C, for the best accuracy constant temperature of the photometer and the reagents is recommended. Air-conditioner recommended (option)
|
|
Power Supply:
|
125 – 240 VAC, 50 – 60 Hz, 400 VA, to be specified for installation
|
|
Sampling:
|
Sampling system with fast loop bypass, pre-filtering and pressure-less sampling point (Option).
|
|
Dimensions:
|
62 x 76 x 50 cm (w x h x d)
|
|
Weight:
|
Approx. 80 kg (depending on installed options)
|
|
Provided with delivery:
|
Hg-MAT1 NE in air-conditioned stainless-steel-cabinet, including photometer, state-of-the-art operating software and bins for reagents.
|